4.3 Checking the pressures
CAUTION
Compressed air connection on devices.
Dirty and moist compressed air causes premature bearing wear.
Ensure that the cooling air is dry, clean and uncontaminated in accordance
with EN ISO 7494-2.
Insert the test gauge (Mat. no. 0.411.8731) between the
MULTIflex (LUX) / MULTIflex LED coupling and the medical device, and check
the following pressures:
- Drive pressure: 2.8 bar (40 psi)
- Return air: < 0.5 bar (7 psi)
- Water: 0.8 to 2.5 bar (12 to 36 psi)
4.4 Checking the O-rings
CAUTION
Missing or damaged O-rings.
Malfunctions and premature failure.
Make sure that all O-rings are on the coupling and undamaged.
Number of available O-rings: 5
Instructions for use SMARTtorque LUX S619 L - 1.008.1641 SMARTtorque Mini LUX S615 L - 1.008.1643
SMARTtorque S619 C - 1.008.1642
4 Start up and shut down | 4.3 Checking the pressures
8/21